Characterisation and deployment of an immobilised pH sensor spot towards surface ocean pH measurements.

Clarke, Jennifer S., Achterberg, Eric P. , Rerolle, Victoire M.C., Abi Kaed Bey, Samer, Floquet, Cedric F.A. and Mowlem, Matthew C. (2015) Characterisation and deployment of an immobilised pH sensor spot towards surface ocean pH measurements. Open Access Analytica Chimica Acta, 897 . pp. 69-80. DOI 10.1016/j.aca.2015.09.026.

[thumbnail of Clarke.pdf]
Preview
Text
Clarke.pdf - Published Version
Available under License Creative Commons: Attribution 3.0.

Download (1MB) | Preview

Supplementary data:

Abstract

The oceans are a major sink for anthropogenic atmospheric carbon dioxide, and the uptake causes changes to the marine carbonate system and has wide ranging effects on flora and fauna. It is crucial to develop analytical systems that allow us to follow the increase in oceanic pCO(2) and corresponding reduction in pH. Miniaturised sensor systems using immobilised fluorescence indicator spots are attractive for this purpose because of their simple design and low power requirements. The technology is increasingly used for oceanic dissolved oxygen measurements.

We present a detailed method on the use of immobilised fluorescence indicator spots to determine pH in ocean waters across the pH range 7.6-8.2. We characterised temperature (-0.046 pH/degrees C from 5 to 25 degrees C) and salinity dependences (-0.01 pH/psu over 5-35), and performed a preliminary investigation into the influence of chlorophyll on the pH measurement. The apparent pK(a) of the sensor spots was 6.93 at 20 degrees C. A drift of 0.00014 R (ca. 0.0004 pH, at 25 degrees C, salinity 35) was observed over a 3 day period in a laboratory based drift experiment. We achieved a precision of 0.0074 pH units, and observed a drift of 0.06 pH units during a test deployment of 5 week duration in the Southern Ocean as an under way surface ocean sensor, which was corrected for using certified reference materials. The temperature and salinity dependences were accounted for with the algorithm, R = (0.00034 - 0.17.pH + 0.15.S-2 + 0.0067.T - 0.0084.S) . 1.075. This study provides a first step towards a pH optode system suitable for autonomous deployment. The use of a short duration low power illumination (LED current 0.2 mA, 5 mu s illumination time) improved the lifetime and precision of the spot. Further improvements to the pH indicator spot operations include regular application of certified reference materials for drift correction and cross-calibration against a spectrophotometric pH system. Desirable future developments should involve novel fluorescence spots with improved response time and apparent pK(a) values closer to the pH of surface ocean waters.

Document Type: Article
Additional Information: WOS:000363419800009
Keywords: Seawater; pH sensor; Fluorescence; Optode; Ocean acidification; JR274
Research affiliation: OceanRep > GEOMAR > FB2 Marine Biogeochemistry > FB2-CH Chemical Oceanography
OceanRep > The Future Ocean - Cluster of Excellence
OceanRep > GEOMAR > FB2 Marine Biogeochemistry > FB2-CH Chemical Oceanography > FB2-CH Water column biogeochemistry
Kiel University
Refereed: Yes
Open Access Journal?: No
Publisher: Elsevier
Projects: Future Ocean
Date Deposited: 24 Nov 2015 07:44
Last Modified: 23 Apr 2021 08:54
URI: https://oceanrep.geomar.de/id/eprint/30338

Actions (login required)

View Item View Item