OceanRep
Design of an automated flow injection-chemiluminescence instrument incorporating a miniature photomultiplier tube for monitoring picomolar concentrations of iron in seawater.
Bowie, A. R., Achterberg, Eric P. , Ussher, S. and Worsfold, P. J.
(2005)
Design of an automated flow injection-chemiluminescence instrument incorporating a miniature photomultiplier tube for monitoring picomolar concentrations of iron in seawater.
Journal of Automated Methods and Management in Chemistry, 2005
(2).
pp. 37-43.
DOI 10.1155/JAMMC.2005.37.
Preview |
Text
JAMMC-2005-037.pdf - Published Version Download (1MB) | Preview |
Abstract
A flow-injection (FI)-based instrument under Lab VIEW control for monitoring iron in marine waters is described. The instrument incorporates a miniature, low-power photomultiplier tube (PMT), and a number of microelectric and solenoid actuated valves and peristaltic pumps. The software allows full control of all flow injection components and processing of the data from the PMT. The optimised system is capable of 20 injections per hour, including preconcentration and wash steps. The detection limit (3 sd of the blank) is 21 pM at sea and the linear range is 21-2000 pM with a 60-second sample load time. Typical precision between replicate FI peaks is 5.9 ± 3.2 (n = 4) over the linear range. © 2005 Hindawi Publishing Corporation.
Document Type: | Article |
---|---|
Refereed: | Yes |
Open Access Journal?: | No |
Publisher: | Hindawi Publ. Corp. |
Date Deposited: | 03 Feb 2014 13:49 |
Last Modified: | 27 Jul 2020 08:13 |
URI: | https://oceanrep.geomar.de/id/eprint/21793 |
Actions (login required)
![]() |